Description Of The Error Message System; Procedure While Error 14, From Software Version 1.76 - Hermle Z 446 K User Manual

Table of Contents

Advertisement

TROUBLE SHOOTING

4.2.2 Description of the Error Message System

The error message, "error" (M11), is shown in the "time" (A-3) display, (see Figure 34). For more
detailed information, refer to Table 6: "Error Messages", (see Appendix P.VIII).
Figure 34

4.2.3 Procedure while error 14, from software version 1.76

If Error 14 occurs, there is a problem with the speed sensor. The centrifuge lid is closed for
undefined period of time and in the "rpm/rcf" (A-1) display shows the lettering "USEr GuidE" (see
Figure 35).
Figure 35
To reopen the centrifuge lid, switch off the device and wait until the rotor has come to a standstill.
Take from "
the respective rotor. Level 0 corresponds to unbreaked rundown, which occurs at error 14. If the
centrifuge lid is opened before standstill of the rotor, a following error can occur.
Once the rotor has come to a standstill, open the centrifuge lid with the emergency release. Proceed
as described in chapter 4.2.1, p 26. After opening the centrifuge lid, switch on the device again.
Error 14 and the lettering "USEr GuidE" should be eliminated.
34
Table 5: Acceleration and Deceleration Times
Z446K_V1.22_eng
" the maximum deceleration time of
© Hermle Labortechnik GmbH

Advertisement

Table of Contents
loading

Table of Contents