Touch Current Measurement; Detergents And Disinfectants - Miele PW 6 7 Series Technical Information

Table of Contents

Advertisement

Technical Information
The ground connection is made only if the ground screws are screwed into
their correct positions during reassembly.
When dismantling the front and side panels, mark the positions of the ground
screws.
During reassembly, screw the ground screws into these same positions.
To check the proper screw positions, measure the resistance of the ground
connection loop.
6

Touch Current Measurement

Note:
A touch current measurement has to be carried out on all accessible
conductive parts that are not connected to ground.
Warning!
Touch current measurement should only be carried out after the ground
connection of the machine under test has been checked and found to be
satisfactory!
Dangerous voltages may exist on defective appliances as well as on
accessible conductive parts that are not connected to ground!
Note:
Do a touch current measurement on the following accessible conductive
parts:
Door ring.
7

Detergents and Disinfectants

Danger!
The warnings and instructions on detergent and disinfectant containers
must be followed closely. The operator is responsible for ensuring that
all containers are clearly marked with suitable warnings when detergents
and disinfectants are transferred from large containers into smaller ones for
use at the installation site.
8
PW 6xx7