Setting Periodic Calibration Reminders - Konica Minolta CS-3000HDR Instruction Manual

Spectroradiometer
Table of Contents

Advertisement

Setting Periodic Calibration Reminders

Periodic calibration (once a year) maintains high measurement accuracy.
When the periodic calibration deadline approaches, a caution message reminding the user of the
periodic calibration can be displayed at the startup of the instrument. If the reminder setting is set to
[ON], a warning message will be displayed at startup when 11 months have passed since the starting
point date.
* Factory default setting:
At first startup, a screen display will appear to select the reminder setting [ON] or [OFF]. If skipped, the
setting will be set to [OFF].
* The starting point date is set as the calibration date, service date, first startup date, etc.
Operating Procedure
1
1
. Press MENU key when the MEAS screen
is displayed.
The MENU screen appears on the LCD.
When the backlight of the LCD has been turned off via
BACKLIGHT
backlight is turned on.
2
. Press either
[SETUP] and then press ENTER key.
The MENU - SETUP screen appears on the LCD.
6, 7
key on the MEAS screen, the
or
key to select
M E A S
Lv
x
y
CMF
MEAS
MEMORY
O P T I O N
S E T U P
D A R K M E A S U R E M E N T
I N F O R M A T I O N
64
2, 3, 4, 5
S N G L
U C 0 0
c d
m
2
SPD
SYN
[
Hz
]
ACC
Nr m
5 9 . 9 4
Non
M E N U
M E N U
S E T U P
(
3/3
)
D A T E F O R M A T
Y Y Y Y / M M / D D
C A L R E M I N D I N G
O F F