Konica Minolta CS-3000HDR Instruction Manual page 69

Spectroradiometer
Table of Contents

Advertisement

2
. Press either
[OPTION] and then press ENTER key.
The MENU - OPTION screen appears on the LCD.
The current set contents are displayed in the [USER
CAL] item.
3
. Press either
CAL] and then press ENTER key.
The MENU - OPTION - USER CAL (for selection of
calibration channel) screen appears on the LCD.
The calibration channel number and compensation
coefficient ID (maximum of 10 characters) are
displayed. In the case of Ch00, "NON" is displayed.
4
. Press either
channel.
Press
key for a larger number.
Press
key for a smaller number.
The calibration channel selectable range is OFF and 01 to 10.
5
. Press ENTER key.
When the calibration channel is set, the MENU -
OPTION screen appears again on the LCD.
If a calibration channel without set correction factor is
selected, it cannot be set.
By pressing the ESC key, the setting is canceled and
the MENU - OPTION screen appears again on the LCD.
The calibration channel setting is saved even after the
power switch is turned OFF (O).
6
. Press ESC key.
The MENU screen appears again on the LCD.
7
. Press ESC key.
The MEAS screen appears again on the LCD.
or
key to select
or
key to select [USER
or
key to select a
67
M E N U
O P T I O N
(
1/2
)
N E G A T I V E V A L U E
N O P R O C
U S E R C A L
O F F
C L O S E U P
O F F
E X T - N D
O F F
M E N U
(
)
O P T I O N
1/2
U S E R C A L
   CAL NO.
    OFF
M E N U
O P T I O N
(
1/2
)
U S E R C A L
   CAL NO.
    03
  AB-X
M E N U
O P T I O N
(
1/2
)
N E G A T I V E V A L U E
N O P R O C
U S E R C A L
0 3
C L O S E U P
O F F
E X T - N D
O F F
M E A S
S N G L
U C 0 3
Lv
63.60
c d
x
0 . 4015
y
0 . 4061
CMF
SPD
SYN
[
Hz
]
ACC
Nrm
5 9 . 9 4
Non
m
2