Konica Minolta CS-3000HDR Instruction Manual page 105

Spectroradiometer
Table of Contents

Advertisement

Model
Measurement
wavelength range
Wavelength resolution
Display wavelength interval
Wavelength precision
±0.3 nm (center-of-gravity wavelength mercury cadmium lamp: 435.8 nm, 546.1 nm, 643.8 nm)
Spectrum wavelength width
Measurement angle
(electrically switchable)
Luminance range with
guaranteed accuracy
(Light source A)
Minimum
ø5 mm (ø1 mm when using
measurement diameter
Minimum objective distance
Minimum luminance display
Minimum spectral
radiance display
Luminance: Accuracy
*1
(Light source A)
1.5%
Luminance:
0.7%
Repeatability
(2σ)
*2
0.25% (0.003 to 0.05 cd/m
(Light source A)
0.15% (0.05 to 5,000 cd/m
x: ±0.002,
Chromaticity:
x: ±0.0015, y: ±0.001 (0.05 to 5,000 cd/m
Accuracy
*1
u': ±0.0022, v': ±0.0011 (0.001 to 0.05 cd/m
(Light source A)
u': ±0.0014, v': ±0.0006 (0.05 to 5,000 cd/m
x: 0.0030, y: 0.0035 (0.001 to 0.003 cd/m
x: 0.0010, y: 0.0015 (0.003 to 0.1 cd/m
x: 0.0006, y: 0.0006 (0.1 to 0.2 cd/m
Chromaticity:
Repeatability
x: 0.0004, y: 0.0004 (0.2 to 5,000 cd/m
(2σ)
*2
u': 0.0024, v': 0.0014 (0.001 to 0.003 cd/m
(Light source A)
u': 0.0009, v': 0.0006 (0.003 to 0.1 cd/m
u': 0.0005, v': 0.0002 (0.1 to 0.2 cd/m
u': 0.0003, v': 0.0002 (0.2 to 5,000 cd/m
Polarization error
2% or less (400 to 780 nm): 1° and 3% or less (400 to 780 nm): 0.2° and 0.1°
Integration time
0.005 to 92 seconds (NORMAL mode); 0.005 to 16 seconds (FAST mode)
Minimum of 1 second or less (MANUAL mode) to approx. 190 seconds (NORMAL mode)
Measurement
Standalone
or maximum of approx. 242 seconds (MANUAL mode)
time
Communication
*3
Approx. 0.07 seconds (when MANUAL mode, 33.333 ms and INTELLIGENT DARK function are set)
Color space mode
L
x y, L
v
Color matching
CIE1931 (2°), CIE1964 (10°), CIE170-2:2015 (PA2°, PA10°), any isochromatic function (measurement
functions
software supported)
Optical frequency
Detectable range: Luminance of 10 to 5,000 cd/m
detection function
Detection accuracy: ±0.015 Hz; Detection time: Approx. 3 seconds
Interfaces
USB 2.0, RS-232C
Operating
temperature and
5 to 30°C, 80% RH max. (no condensation)
humidity range
Storage temperature
0 to 35°C, 80% RH max. (no condensation)
and humidity range
Power
Dedicated AC adapter (100 to 240 V , 50/60 Hz)
Power consumption
Approx. 20 W
Size
158 (W) × 262 (H) × 392 (D) mm
Weight
Approx. 7.0 kg
*1: Average value of 10 measurements in the NORMAL mode, at temperature of 23°C ± 2°C and maximum relative humidity of 65%.
*2: Average value of 10 measurements in the NORMAL mode, at temperature of 23°C ± 2°C and maximum relative humidity of 65%.
*3: When the USB cable is connected. Excluding the time required for display and shutter opening/closing. In an environment
designated by KONICA MINOLTA such as a PC.
CS-3000 SPECTRORADIOMETER
0.0005 to 5,000 cd/m
2
closeup lens)
350 mm (55 mm when using closeup lens)
(0.0005 to 0.001 cd/m
2
)
1.5%
(0.001 to 0.003 cd/m
2
)
0.7%
)
0.25% (0.075 to 1.25 cd/m
2
2
)
0.15% (1.25 to 125,000 cd/m
y: ±0.002 (0.001 to 0.05 cd/m
2
)
x: ±0.002,
2
)
x: ±0.0015, y: ±0.001 (1.25 to 125,000 cd/m
2
)
u': ±0.0022, v': ±0.0011 (0.025 to 1.25 cd/m
2
)
u': ±0.0014, v': ±0.0006 (1.25 to 125,000 cd/m
2
)
x: 0.0030, y: 0.0035 (0.025 to 0.075 cd/m
2
)
x: 0.0010, y: 0.0015 (0.075 to 2.5 cd/m
2
)
x: 0.0006, y: 0.0006 (2.5 to 5 cd/m
2
)
x: 0.0004, y: 0.0004 (5 to 125,000 cd/m
2
)
u': 0.0024, v': 0.0014 (0.025 to 0.075 cd/m
2
)
u': 0.0009, v': 0.0006 (0.075 to 2.5 cd/m
2
)
u': 0.0005, v': 0.0002 (2.5 to 5 cd/m
2
)
u': 0.0003, v': 0.0002 (5 to 125,000 cd/m
u' v', L
TΔuv, XYZ, spectrograph, main wavelength, excitation purity
v
v
103
380 to 780 nm
0.9 nm/pixel
1.0 nm
5 nm max. (half width value)
0.2°
0.0125 to 125,000 cd/m
2
ø1 mm (ø0.2 mm when using
closeup lens)
0.00002 cd/m
2
1.0 × 10
–9
W/(sr·m
2
·nm)
±2%
(0.0125 to 0.025 cd/m
2
)
(0.025 to 0.075 cd/m
2
)
)
2
2
)
y: ±0.002 (0.025 to 1.25 cd/m
2
)
2
)
2
)
2
)
2
)
2
)
2
)
2
)
2
)
2
)
2
)
2
)
2
and emission frequency of 10 to 200 Hz.
0.1°
0.05 to 500,000 cd/m
2
ø0.5 mm (ø0.1 mm when using
closeup lens)
1.5%
(0.05 to 0.1 cd/m
2
)
0.7%
(0.1 to 0.3 cd/m
2
)
0.25% (0.3 to 5 cd/m
)
2
0.15% (5 to 500,000 cd/m
2
)
x: ±0.002,
y: ±0.002 (0.1 to 5 cd/m
2
)
x: ±0.0015, y: ±0.001 (5 to 500,000 cd/m
2
)
u': ±0.0022, v': ±0.0011 (0.1 to 5 cd/m
2
)
u': ±0.0014, v': ±0.0006 (5 to 500,000 cd/m
2
)
x: 0.0030, y: 0.0035 (0.1 to 0.3 cd/m
2
)
x: 0.0010, y: 0.0015 (0.3 to 10 cd/m
2
)
x: 0.0006, y: 0.0006 (10 to 20 cd/m
2
)
x: 0.0004, y: 0.0004 (20 to 500,000 cd/m
u': 0.0024, v': 0.0014 (0.1 to 0.3 cd/m
2
)
u': 0.0009, v': 0.0006 (0.3 to 10 cd/m
2
)
u': 0.0005, v': 0.0002 (10 to 20 cd/m
2
)
u': 0.0003, v': 0.0002 (20 to 500,000 cd/m
2
)
2
)

Advertisement

Table of Contents
loading

This manual is also suitable for:

Cs-3000Cs-2000plus

Table of Contents