Selecting Measurement Angle

The measurement angle can be selected from 1°, 0.2°, or 0.1°.
For the CS-3000HDR/CS-3000, the measurement angle is switched electrically according to the selec-
tion made on the MENU screen. Motor noise may be heard when switching the measurement angle, but
this is not abnormal.
For the CS-2000Plus, the measurement angle is switched by manually rotating the measurement angle
selector.
For details on measurement distance and diameters, refer to the below table. Attach the closeup lens
(optional accessory) if necessary.
* Factory default setting: 1°
Minimum
measurement
diameter ø
(Measurement
angle) 1°
0.2° 0.1°
Without
5.00 1.00 0.50
closeup lens
With closeup
1.00 0.20 0.10 1.39 0.28 0.14
lens
* The measurement distance is the distance from the tip of the objective lens or the metallic part of the closeup lens.
Operating Procedure
1
1
. Press MENU key when the MEAS screen
is displayed.
The MENU screen appears on the LCD.
When the backlight of the LCD has been turned off via
BACKLIGHT
backlight is turned on.
Maximum
Minimum
measurement
objective
diameter ø
distance
0.2° 0.1°
0.2° 0.1°
350
55.0
5, 6
key on the MEAS screen, the
Measurement
Maximum
diameter ø for 500
objective
mm measurement
distance
0.2° 0.1°
7.78 1.56 0.78 16.66 3.33 1.67
70.9
-
M E A S
Lv
x
y
CMF
MEAS
MEMORY
O P T I O N
S E T U P
D A R K M E A S U R E M E N T
I N F O R M A T I O N
40
(Unit: mm)
Measurement
diameter ø for 1,000
mm measurement
distance
distance
0.2° 0.1°
0.2° 0.1°
-
-
-
-
-
2, 3, 4
S N G L
U C 0 0
c d
m
2
SPD
SYN
[
Hz
]
ACC
Nr m
5 9 . 9 4
Non
M E N U