Selecting Synchronous Frames

In synchronized measurement, synchronous frames can be selected from "SINGLE FRAME" or "DOUBLE
FRAME."
In the case of "SINGLE FRAME," synchronized measurement is performed with the vertical scan signal
period as one cycle.
In the case of "DOUBLE FRAME," synchronized measurement is performed with one cycle being twice
the vertical scan signal cycle.
When flickering occurs on the LCD, the screen is fluctuating at half the vertical scanning frequency. For
stable measurement of LCDs, it is recommended to set the integration time to twice the vertical scan
period ("DOUBLE FRAME").
* Factory default setting: DOUBLE FRAME
Operating Procedure
1
1
. Press MENU key when the MEAS screen
is displayed.
The MENU screen appears on the LCD.
When the backlight of the LCD has been turned off via
BACKLIGHT
backlight is turned on.
2
. Press either
and then press ENTER key.
The MENU - MEAS screen appears on the LCD.
The current set contents are displayed in the [SYNC
FRAME] item.
6, 7
key on the MEAS screen, the
or
key to select [MEAS]
M E A S
Lv
x
y
CMF
MEAS
MEMORY
O P T I O N
S E T U P
D A R K M E A S U R E M E N T
I N F O R M A T I O N
38
2, 3, 4, 5
S N G L
U C 0 0
c d
m
2
SPD
SYN
[
Hz
]
ACC
Nr m
5 9 . 9 4
Non
M E N U
M E N U
M E A S
S P E E D
NORMAL
S Y N C
INT SYNC
S Y N C F R A M E
DOUBLE FRAME
A N G L E

Advertisement

Table of Contents
loading

This manual is also suitable for:

Cs-3000Cs-2000plus

Table of Contents