which may break the instrument. Also, use the instrument in a well-ventilated place. To ensure
proper heat dissipation, keep the ventilation holes free from obstruction.
 Avoid a rapid change in ambient temperature to prevent condensation.
 Avoid using the instrument in extremely dusty or humid places.
 Use the CS-2000Plus at an ambient temperature between 5 and 35ºC and relative humidity of 80%
or less (at 35ºC) with no condensation. Use the CS-3000HDR/CS-3000 at an ambient temperature
between 5 and 30ºC and relative humidity of 80% or less (at 30ºC) with no condensation. Operating
the instrument outside the specified temperature and humidity ranges may impede its performance.
 Do not use the instrument at altitudes higher than 2,000 m above sea level.
 Make sure the AC adapter output plug is not short-circuited. A short-circuit may cause fire or electric shock.
 Do not connect the AC adapter to an overloaded electrical circuit. In addition, do not cover or wrap
the AC adapter with cloth or any other material while in use. Doing so may cause electric shock or fire.
 When removing the AC adapter from the instrument, first remove the power cord from the outlet,
and then remove the output plug.

This Instrument

 Do not subject the instrument to strong impact or vibration.
 Do not forcibly pull, bend, or apply strong force to the power cord for the included AC adapter or
USB cable. This may result in the cord snapping.
 Connect the instrument to a power source with minimal noise.
 Do not measure a high-luminance light source (including sunlight) beyond the measurement range.
Failure to observe this warning could result in damage to the instrument's optical system.
 Should you notice any breakage or abnormality during operation, immediately switch the power off
and unplug the AC adapter. Then refer to "Error Check." p.98
 Should this instrument break down, do not try to disassemble and repair it by yourself. Please
contact the nearest KONICA MINOLTA-authorized service facility.
 Warm this instrument up for at least 20 minutes after switching the power on when the object
luminance is 2 cd/m2or lower (measurement angle 1º).
 When not using RS-232C communication, be sure to attach the connector cap. Failure to do so may
cause malfunction due to static electricity.
Objective Lens, ND Filter, Closeup Lens, and Illuminance Adapter (Optional Accessories)
 When performing measurements, make sure that the surfaces of the objective lens, ND filter,
closeup lens, and illuminance adapter (optional accessories) are clean. Correct measurement may
not be performed if there is dirt, dust, fingerprints or parts left unclean.
 Do not touch the surface of objective lens, ND filter, closeup lens or illuminance adapter with your hands.
 When a sudden change in temperature is applied in a high humidity environment, this may mist the
objective lens, ND filter, closeup lens or illuminance adapter, resulting in incorrect measurements.
 Please note that observing light of about 100,000 lx with a light source with a large infrared light
output, such as an A light source, may cause a large temperature rise inside the illuminance adapter
and the main unit, resulting in damage.