Time To Spill (2.8.12.); Discrete Inputs (2.9.); Backup Level Override (2.9.1) - Siemens SITRANS L 7ML5400 Operating Instructions Manual

Ultrasonic level controllers
Table of Contents

Advertisement

9.2.8.12

Time To Spill (2.8.12.)

Used to predict when an overflow (spill) condition may occur. This feature works in
conjunction with the High Level Alarm (2.8.1.) (Page 200).
Level To Spill (2.8.12.1.)
Value (defined in Units (2.1.1.) (Page 172)) representing material level at which a spill will
occur.
Options
This value must be greater than high level alarm ON setpoint [High Level Value ON (2.8.1.2)
(Page 200) ].
Minutes Left To Spill (2.8.12.2.)
Read only. Calculated value representing minutes remaining before a spill will occur.
Enter the level at which a spill condition will occur in Level To Spill (2.8.12.1.) (Page 211).
When the High Level Alarm is tripped, the estimated time to spill is displayed in
2.8.12.2.Minutes Left To Spill. The estimated time is calculated by the LUT400 based on the
material level and the rate of change of the material level. If the High Level alarm is not
tripped, or the material level is falling, then the estimated time to spill will display as zero.
9.2.9

Discrete Inputs (2.9.)

Discrete inputs are used to trigger or alter the way SITRANS LUT400 controls devices such as
pumps and alarms. Discrete inputs can be used for the following:
• as a backup level override
• allowing the device to be more flexible by interlocking control functions with external
conditions.
For more detail see Discrete Inputs (Page 91).
Note
Default settings in the parameter tables are indicated with an asterisk (*) unless explicitly
stated.
9.2.9.1

Backup Level Override (2.9.1)

Use this feature to override the material reading by a discrete input such as a contacting point
device. The material reading will be fixed at the programmed switch level until the discrete
input is released. The LUT400 makes decisions based on the override values.
SITRANS LUT400
Operating Instructions, 04/2021, A5E33329501-AG
Parameter reference (LUI)
-999999.000 to 999999.000
Default: 0.000
9.2 Setup (2.)
211