OPERATION

2.3.4 Pre-Selection of Temperature

This function is activated by the key, "temp" (13). After pressing the key, in the display "temp", flashes
the indication, "°C" (A-4). With the adjustable knob (1), the desired test temperature can be pre-
selected, in steps of 1°C in a range from: -20°C to +40°C.
The value is indicated permanently in the display, (Figure 21) - before, during and after the run.
Please notice the respective lowest temperatures of the rotors, at maximum speed!
Figure 21

2.3.5 Pre-Cooling

If the samples are "temperature sensitive", it is critical to pre-cool the centrifuge, the rotor and
eventually the buckets, to the needed working temperature. Therefore, insert the desired rotor and
pre-set the respective temperature. By simultaneously pressing the keys, "temp" (13) and "time" (6),
the run will begin. While the unit is running, it automatically choose a rotational speed that is
equivalent to 20 % of the permitted rotational speed, of the respective rotor. After the pre-set
temperature is reached, you can opt out of the pre-cooling run, with the "stop" key (10).
Depending on the inserted rotor, the pre-cooling lasts between 10 to 20 min.
For all number marked text, please refer to Figure 22.
Figure 22
24