14. Retrofitting Components

14.1
Fitting the Filter Magazine
(Transmitted Light)
• Remove the tube and intermediate systems
where applicable.
• Turn the microscope stand upside down,
loosen the fastening screws at the bottom and
lift off the base plate.
• Insert the filters into the semicircular mounts.
This does not have to be done in any particu-
lar order.
• Put the filter magazine back in position.
Fig. 71
Filter magazine (transmitted light) for Leica DM2500
14. Retrofitting Components
14.2
Equipping the Condenser Disk
• Turn the stage upward and lower the con-
denser.
• Remove the condenser. Therefore loosen the
condenser's clamping screw.
Condenser UCL/UCLP
• Remove the screw (72.1) completely.
• Turn back the centering screws until the light
rings, λ - and λ/4 - compensator* and lens*
2.5x can be inserted.
The largest hole is for brightfield observa-
tion (= BF), the slightly smaller ones for light
rings or λ - and λ/4 - compensator or lens*
2.5x.
Fig. 72
Condenser UCL
1
Fixing screw for condenser disk
1
67