Preparation Before Surgery
2
1
Adjust the tilt of the footswitch. Rotate the footrest in or out.
¸
¸
Check that all accessories are correctly connected and
¸
¸
securely attached.
Release the articulation brakes and set them to a light
¸
¸
resistance.
Make the joint easier to move:
Release the articulation brake (1).
¸
¸
Make the joint more difficult to move:
Tighten the articulation brake (1).
¸
¸
Pull the retaining lever (2).
¸
¸
Extend the swing arm.
¸
¸
Check the weight setting on the swing arm by raising and
¸
¸
lowering the microscope and correct as necessary, see
page 8.
Caution
Risk of infection!
Leave sufficient space around the stand to prevent
¸
¸
contact with non-sterile components.
16
1
Leica M620 F20 / Ref. 10 714 371 / Version 00

Installing sterile components

1
Standard components:
• One handle, large, black (tilt head)
• 2 handles, transparent
• 2 drive knobs
The handles are steam and gas-sterilizable.
Sterilize the handles and drive knobs.
¸
¸
Attach the sterile drive knobs on the binocular tube and the
¸
¸
drive knob to block ambient illumination.
Attach sterile handles to optics carrier and tilt head.
¸
¸

Checking the function of the lamp

Switch on the microscope using the power switch.
¸
¸
The main lamp will light up.
Move the slide button for the quick-change lamp mount (3)
¸
¸
to the other side.
The auxiliary lamp will light up.
Move the slide button for the quick-change lamp back to the
¸
¸
main lamp.
The slide button for the quick-change lamp mount (3)
must be slid fully against its stop, otherwise the lamp
will not light.
3