Leica DM2000 Operating Manual page 43

Hide thumbs Also See for DM2000:
Table of Contents

Advertisement

Available languages

Available languages

Note:
Close any unused tube openings, as otherwise
stray light can interfere with observation.
Adjusting the Viewing Distance
• Adjust the viewing distances of the eye-
pieces so that a congruent total image is
seen (Fig. 52).
Adjusting the Viewing Angle
• For the HC LVB 0/4/4 and HC -/0/4 ergonomy
tubes, the viewing angle can be adjusted by
tilting the binocular viewer.
Ergotube (long, swivelable):
Ergotube (short, swivelable):
• For the AET22 and EDT22 ergotubes, the view-
ing angle can be adjusted by tilting the bin-
ocular viewer in the range of 5° - 32° (Fig. 53).
Fig. 52 Tube setting
↔ Personal eyebase settings
1 Scale (mm), 2 Intermediate module*, in illustration: Ergo
module
0° - 35°
7.5° - 32.5°
1
2
Adjusting the Eyepiece Section to the Arm
Length
• On the AET22 tube, the eyepieces can be
extended up to 30 mm (Fig. 53).
Beam Splitting in Photo Tubes
EDT22 tube:
The beam splitting between the observation and
documentation outputs has a definite presetting
(50%:50%).
BDT25+ tube:
The beam splitting is set manually by pulling out
a control bar.
Control Bar
VIS
50/50
PHOTO
Fig. 53
With AET22 tube individual adjustments
8. Operation
Observation
Photo
100%
150%
110%
100%
0%
50%
43

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Dm2500

Table of Contents