Reference Manual
00809-0100-4026, Rev KC
Le présent appareil est conforme aux CNR d'Industrie
Canada applicables aux appareils radio exempts de licence.
L'exploitation est autorisée aux deux conditions suivantes:
(1) l'appareil ne doit pas produire de brouillage, et (2)
l'utilisateur de l'appareil doit accepter tout brouillage
radioélectrique subi, même si le brouillage est susceptible
d'en compromettre le fonctionnement.
Certificate: 2827A-5401
2827A-5402
B.3.3 Radio Equipment Directive
(RED)
This device complies with ETSI EN 302 372 and EN 62479.
EU directive 2014/53/EU.
B.4 Installing Equipment in
North America
The US National Electrical Code (NEC) and the Canadian
Electrical Code (CEC) permit the use of Division marked
equipment in Zones and Zone marked equipment in
Divisions. The markings must be suitable for the area
classification, gas, and temperature class. This information
is clearly defined in the respective codes.
B.5 USA
E5 Explosion-proof (XP), Dust- ignition-proof (DIP)
Certificate: FM16US0444X
Standards: FM Class 3600 – 2011; FM Class 3610 –
2010;
FM Class 3611 – 2004; FM Class 3615 – 2006;
FM Class 3810 – 2005; ANSI/ISA 60079-0 –
2013; ANSI/ISA 60079-11 – 2012;
ANSI/NEMA 250 – 2003
Markings: XP CL I, DIV 1, GP B, C, D; DIP CLII/III, DIV 1,
GP E, F, G; T4 Ta=60°C and 70°C; Type 4X
Product Certifications
Specific Conditions for Safe Use (X):
1. WARNING - Potential Electrostatic Charging Hazard –
The enclosure contains non-metallic material. To
prevent the risk for electrostatic sparking, the plastic
surface should only be cleaned with a damp cloth.
2. WARNING - The apparatus enclosure contains
aluminum and is considered to constitute a potential
risk of ignition by impact or friction. Care must be
taken into account during installation and use to
prevent impact or friction.
I5
Intrinsic Safety (IS), Nonincendive (NI)
Certificate: FM16US0444X
Standards: FM Class 3600 – 2011; FM Class 3610 –
2010;
FM Class 3611 – 2004; FM Class 3615 – 2006;
FM Class 3810 – 2005; ANSI/ISA 60079-0 –
2013;
ANSI/ISA 60079-11 – 2012;
ANSI/NEMA 250 – 2003;
Markings: IS CL I, II, III, DIV 1, GP A, B, C, D, E, F, G in
accordance with control drawing
9150079-905;
IS (Entity) CL I, Zone 0, AEx ia IIC T4 in
accordance with control drawing
9150079-905, NI CL I, II,
DIV 2, GP A, B, C, D, F, G;
Suitable for use in CL III DIV 2, indoor and
outdoor, T4 Ta=60 °C and 70°C; Type 4X
Specific Conditions for Safe Use (X):
1. WARNING - Potential Electrostatic Charging Hazard –
The enclosure contains non-metallic material. To
prevent the risk for electrostatic sparking, the plastic
surface should only be cleaned with a damp cloth.
2. WARNING -The apparatus enclosure contains
aluminum and is considered to constitute a potential
risk of ignition by impact or friction. Care must be
taken into account during installation and use to
prevent impact or friction.
U
i
Entity parameters
30 V
HART
Entity parameters
30 V
Fieldbus
Product Certifications
October 2018
I
P
C
i
i
i
130 mA
1 W
7.26 nF
300 mA
1.3 W
0
L
i
0
0
242