7 Menu Functions
Channel Menu
and humidity values displayed are averaged over the measurement period.
When average is off, the displayed values are instantaneous measurements.
RATE T, rate time, determines the time interval over which the measurements
are used to calculate the rate of change, and is selectable between 30 seconds,
1, 2, 5, 10, 15, 20, 30, 45 minutes, or 1 hour. The temperature rate is displayed
in the RATE T field of a STATS display zone. The humidity rate is displayed in
the RATE H field of a STATS display zone. (See Section 7.2.3.) If RATE T is 5
minutes or less, the rate of change is calculated as the average rate of change
using all the measurements obtained during the past RATE T time. If RATE T
is greater than 5 minutes, the rate of change is calculated using the most recent
measurement and the measurement obtained RATE T time ago. If there are as
yet insufficient measurements to calculate the rate of change, only dots are dis-
played in the RATE T or RATE H field.

Figure 8 Channel Setting

Use the ud buttons to select the setting to change. The lr buttons are used
to change the setting parameters. Press the Enter button to save the new set-
ting. Press the Exit button to cancel any changes and exit to the menu.
7.1.2

Sensor ID

The SENSOR ID function allows each sensor to be uniquely identified. This
may be password-protected to prevent it from being changed (see Section 7.5.4,
PASSWORD). The model number and serial number are displayed here but
27