ALARm:TEMPurature<chn>:LOWer:LIMit? [MIN|MAX|DEF]
Returns the lower temperature alarm limit for the specified channel in degrees
C or F.
Example: ALAR:TEMP1:LOW:LIM?
Response: 18.00
The TEMPerature suffix, <chn>, specifies the channel (1 or 2). The value re-
turned is always in the current temperature units (C or F). Specifying the MIN,
MAX, or DEF parameter returns the minimum, maximum, or default value
respectively.
ALARm:TEMPurature<chn>:LOWer:LIMit <float>|MIN|MAX|DEF
Sets the lower temperature alarm limit for the specified channel in degrees C or
F.
Example: ALAR:TEMP1:LOW:LIM 20.00
The TEMPerature suffix, <chn>, specifies the channel (1 or 2). The <float> pa-
rameter specifies the value for the lower temperature alarm limit. Specifying
the MIN, MAX, or DEF parameter sets the lower temperature alarm limit to the
minimum, maximum, or default value respectively. The <float> value is always
in the current temperature units (C or F). This command may be password pro-
tected (see Section 8.2.10, Password Commands).
ALARm:TEMPurature<chn>:RATE?
Indicates whether a temperature rate alarm event has occurred.
Example: ALAR:TEMP1:RATE?
Response: 0
The TEMPerature suffix, <chn>, specifies the channel (1 or 2). A value of 1 is
returned if a temperature rate alarm event has occurred on the specified chan-
nel. A value of 0 is returned if a temperature rate alarm event has not occurred
on the specified channel.
ALARm:TEMPurature<chn>:RATE:ENABle?
Returns the state of the temperature rate alarm enable for the specified channel.
Example: ALAR:TEMP1:RATE:ENAB?
Response: 0
The TEMPerature suffix, <chn>, specifies the channel (1 or 2). A value of 1 is
returned if the temperature rate alarm is enabled. A value of 0 is returned if the
temperature rate alarm is disabled.