1620A "DewK" Thermo-hygrometer
User's Guide
8.2.1
8.2.1.1
8.2.1.2
8.2.1.3
8.2.1.4
8.2.1.5
80
• Unrecognized commands or commands with incorrect syntax or invalid
parameters generate error messages that are placed in the error queue (see
Section 8.2.7.6, SYST:ERR?).

Alarm Commands

ALARm:BATTery?

Indicates whether a battery alarm event has occurred.
Example: ALAR:BATT?
Response: 1
A value of 1 is returned if a battery alarm event has occurred. A value of 0 is
returned if a battery alarm event has not occurred.

ALARm:BATTery:ENABle?

Returns the state of the battery alarm enable.
Example: ALAR:BATT:ENAB?
Response: 0
A value of 1 is returned if the battery alarm is enabled. A value of 0 is returned
if the battery alarm is disabled.
ALARm:BATTery:ENABle <bool>
Sets the state of the battery alarm enable.
Example: ALAR:BATT:ENAB 1
The <bool> parameter turns the battery alarm enable on (1 or ON) or off (0 or
OFF). This command may be password protected (see Section 8.2.10, Pass-
word Commands).

ALARm:BEEP:ENABle?

Returns the state of the audible alarm enable.
Example: ALAR:BEEP:ENAB?
Response: 0
A value of 1 is returned if the alarm beep is enabled. A value of 0 is returned if
the alarm beep is disabled.
ALARm:BEEP:ENABle <bool>
Sets the state of the audible alarm enable.
Example: ALAR:BEEP:ENAB 1