Figure 46 Password - Fluke 1620A-H User Manual

Thermo-hygrometer
Hide thumbs Also See for 1620A-H:
Table of Contents

Advertisement

1620A "DewK" Thermo-hygrometer
User's Guide
66
the ENTER button to accept the password or press the EXIT button to cancel
any changes and exit to the menu.

Figure 46 Password

By default the password is set to "1620" when the instrument ships from the
factory. We recommend the user changes the default password to secure the
password-protected settings.
NOTE: Keep your password in a secure location.
If an incorrect password is entered, the message WRONG PASSWORD is dis-
played and access to the password-protected parameters is denied. Press the
ENTER button to reenter the password.
If a correct password is entered, the password-protected parameters PASS,
SENSOR, RECORD, STATS, ALARM, TIME, and COMM are displayed.
The PASS parameter allows the user to change the password and consists of a
four digit number.
The SENSOR parameter allows the user to turn the password-protection for
sensor and channel settings ON or OFF.
The RECORD parameter allows the user to turn the password-protection for
data recording settings ON or OFF.
The STATS parameter allows the user to turn the password-protection of statis-
tics and daily stats settings ON or OFF.
The ALARM parameter allows the user to turn the password-protection of
alarm settings ON or OFF.
The TIME parameter allows the user to turn the password-protection of time
and date settings ON or OFF.

Advertisement