7 Menu Functions
Channel Menu
desired location. It uses the sensor serial number to identify the sensor. To ex-
change a different sensor, the sensor lock must be set to OFF.

Figure 10 Sensor Lock

Use the ud buttons to select the sensor to lock. The lr buttons are used to
select ON or OFF. Press the Enter button to save the setting and exit. Press the
Exit button to cancel any changes and exit to the menu.
7.1.4

Sensor Cal

The SENSOR CAL function allows the sensor parameters to be edited. This
function is always password-protected (see Section 7.5.4, PASSWORD). If an
incorrect password or no password is entered, the parameters are displayed but
cannot be edited. If a correct password is entered, the channel, sensor model,
sensor serial number, and sensor ID are displayed. Select the sensor channel of
the parameters to edit and then select the Enter button. The sensor parameters
are displayed for editing.
Caution: The sensor calibration parameters are critical to the accuracy of
the sensor. Do not change them (see Section 9, Calibration Procedure).
The lr buttons are used to select a channel. Press the Enter button to save
29