Fluke 1620A-H User Manual page 136

Thermo-hygrometer
Hide thumbs Also See for 1620A-H:
Table of Contents

Advertisement

1620A "DewK" Thermo-hygrometer
User's Guide
124
bit 0 is 1: channel 1 sensor is unattached, locked out, or not operating
properly
bit 1 is 1: channel 2 sensor is unattached, locked out, or not operating
properly
2nd number: channel 1 sensor test
0: no error
bit 0 is 1: sensor unattached
bit 1 is 1: sensor memory read failure
bit 2 is 1: sensor parameter checksum error
bit 3 is 1: sensor calibration parameters out of range
bit 4 is 1: sensor calibration date is invalid
bit 5 is 1: sensor identification is invalid
bit 6 is 1: sensor version number is incompatible
bit 7 is 1: sensor is locked out
3rd number: channel 2 sensor test (see channel 1 sensor test)
4th number: sensor calibration expired
0: sensor calibration has not yet expired
bit 0 is 1: channel 1 sensor calibration has expired
bit 1 is 1: channel 2 sensor calibration has expired
5th number: clock status
0: operating properly
bit 0 is 1: time/date is invalid
bit 1 is 1: clock is not operating
6th number: serial port status
0: operating properly
bit 0 is 1: failure
7th number: RF transceiver status
0: operating properly
bit 0 is 1: failure or not installed
8th number: LAN port status
0: operating properly
bit 0 is 1: failure