Mounting The "Sensor, Remote" Version - Endress+Hauser Levelflex FMP56 Operating Instructions Manual

Foundation fieldbus guided wave radar
Hide thumbs Also See for Levelflex FMP56:
Table of Contents

Advertisement

Mounting
36
Mounting devices with a flange
If a seal is used to mount the device, use uncoated metal screws to ensure good electrical
contact between the process flange and the probe flange.
Mounting rope probes
NOTICE
Electrostatic discharge can damage the electronics.
Ground the housing before lowering the rope probe into the vessel.
*
Pay attention to the following when introducing the rope probe into the vessel:
• Uncoil the rope slowly and lower it carefully into the vessel.
• Make sure the rope does not bend or buckle.
• Avoid uncontrolled swinging of the weight, as this could damage internal fittings in the
vessel.
Mounting rope probes in a partially filled silo
If a silo is retrofitted with a Levelflex, it is not always possible to empty the silo. If the
vessel is at least 2/3 empty, it is possible to mount the rope probe even in a partially
filled silo. In this case, if possible perform a visual inspection after mounting: the rope
should not become tangled or knotted when the silo is emptied. The probe rope must
hang fully extended before an accurate measurement can be performed.
6.2.4

Mounting the "Sensor, remote" version

This section only applies for devices with the version "Probe design" = "Sensor, remote"
(feature 600, version MB/MC/MD).
The following is included in the delivery with the version "Probe design" = "Remote":
• The probe with process connection
• The electronics housing
• The mounting bracket for mounting the electronics housing on a wall or post
• The connection cable (length as ordered). The cable has one straight plug and one plug
angled at 90°. Depending on the local conditions the angled plug can be connected at the
probe or at the electronics housing.
Levelflex FMP56, FMP57 FOUNDATION Fieldbus
*