Operating menu
Medium type
Navigation
Description
User interface
Factory setting
Additional information
Medium property
Navigation
Prerequisite
Description
Selection
Factory setting
144
"Level" submenu
Navigation
Setup → Advanced setup → Level
Setup → Advanced setup → Level → Medium type
Specify type of medium.
• Liquid
• Solid
FMP56, FMP57: Solid
This parameter determines the value of several other parameters and strongly
influences the complete signal evaluation. Therefore, it is strongly recommended not
to change the factory setting.
Setup → Advanced setup → Level → Medium property
EOP level evaluation ≠ Fix DC
Specify the dielectric constant ε
• Unknown
• DC 1.4 ... 1.6
• DC 1.6 ... 1.9
• DC 1.9 ... 2.5
• DC 2.5 ... 4
• DC 4 ... 7
• DC 7 ... 15
• DC > 15
Depends on the Medium type (→ 144) and Medium group parameters.
Levelflex FMP56, FMP57 FOUNDATION Fieldbus
of the medium.
r
Endress+Hauser
Need help?
Do you have a question about the Levelflex FMP56 and is the answer not in the manual?
Questions and answers