Levelflex FMP56, FMP57 FOUNDATION Fieldbus
Endress+Hauser
Diagnostic event and event text
The fault can be identified by means of the diagnostic event. The event text helps you by
providing information about the fault. In addition, the associated status symbol is
displayed in front of the diagnostic event.
Status symbol
Example
If two or more diagnostic events occur simultaneously, only the diagnostic message with
the highest priority is shown. Additional queued diagnostic messages can be shown in the
Diagnostic list submenu.
Past diagnostic messages that are no longer pending are shown as follows:
• On the local display:
In the Event logbook submenu
• In FieldCare:
Via the "Event List /HistoROM" function.
Operating elements
Operating functions in menu, submenu
Plus key
Opens the message about the remedial measures.
Enter key
Opens the operating menu.
Diagnostics and troubleshooting
Diagnostic event
Status signal
Event number
↓
↓
S
441
A0032904
3-digit number
Event text
↓
Curr.output 1
103
Need help?
Do you have a question about the Levelflex FMP56 and is the answer not in the manual?