Installation procedure
Recommended order for
installing end-point
Smart-Vue Dry Contact Module Installation Guide
8
Refer to the Appendix on page 18 for additional considerations.
When routing the cable for any Smart-Vue sensor, avoid direct contact with
or close proximity placement of the sensor cabling with any high voltage
wiring. Cabling should be placed with no less than a minimum of 5 cm (2
inches) distance from high voltage components. Also, avoid running the
sensor cable in parallel with high voltage wiring.
For all other manufacturers' equipment, contact the manufacturer of the
instrument for instructions regarding proper placement of the sensor.
In addition to placing your Smart-Vue end-point module as described above
for best wireless performance, you may also optimize overall network
performance by proceeding with installation as follows:
modules
1. Examine the general layout of your site to determine the placement
of your Smart-Vue end-point modules. The extent of your
pre-installation site survey depends on the number of end-point
modules to install. Feel free to contact your sales representative for
more information on this topic.
2. Place all end-point modules in their final locations before running
wireless setup (described on page 10).
Begin activation by pressing on the end-point module's button for 3 seconds
(long-press), starting with the end-point modules closest to your receiver,
working your way "out" in concentric circles. This enables each Smart-Vue
end-point module to establish the optimal connection to your receiver.
Figure 3. Start installation with end-point modules closest to your
receiver and work your way "out" in concentric circles