1
Installation

Installing the Injector and Detector Modules

Installing the Injector and Detector Modules

Installing an Injector Module

4
TRACE 1300 and TRACE 1310 Hardware Manual
This section provides the instructions for installing your front/back injector and detector
modules into the relevant housing.
The GC is shipped with dummy modules installed into the injector/detector housings.
See
Figure
1.
Figure 1.
Injector and Detector Housings and Dummy Modules
Detector Housing
Note Where a dummy module is installed, the gas connection is blocked by a plug.
The dummy modules must be removed and replaced by the injector and detector modules as
required by the configuration of your TRACE 1300/TRACE 1310.
CAUTION Each housing must always be occupied by both modules. If the GC is
configured with a single injector/detector module, a dummy module must remain
inserted instead of the missing injector/detector module.
The injector and detector modules are shipped with a plug. Before installing the column,
remove the plug.
To install the injector and detector modules, see the following procedures:
Installing an Injector Module
Installing a Detector Module
To install an injector module
1. Remove the dummy module from the position where the injector module will be
installed.
Injector Housing