Software configuration

Additional configuration

Smart-Vue Dry Contact Module Installation Guide
16
Your Smart-Vue end-point module offers many features that are controlled
by your Smart-Vue Server/Smart-Vue Client software. Use the software to
define settings for measurement period, frequency of data transmission,
alarm limits for sending alerts via phone, fax, e-mail and SMS, sensor
calibration/correction parameters, archiving and more. Enter your login
name and password when prompted.
Setting basic sensor parameters
1. Access end-point module settings by pressing F11 or selecting
Setting ==> Sensor settings on the Smart-Vue Client main screen.
Figure 11. Main sensor settings screen for dry contact end-point module
2. Choose a sensor in the left-hand panel (tree structure), and adjust
the specific settings (Sensor Settings tab). You should do this for all
sensors in your system.
Note that the Unit field contains "State". This indicates that the
Smart-Vue end-point module in question is indeed a module for
measuring binary open/closed states on a dry contact device.
We recommend setting both the data transfer and the measurement
intervals to 1 minute so you are notified as quickly as possible if the
monitored device has changed from its normal state.A warning triangle is
shown in the Smart-Vue display if the end-point module's current status is
in not "normal" (i.e., closed instead of open, or open instead of closed).