Kodak X-Omat 2000 Operator's Manual page 19

Hide thumbs Also See for X-Omat 2000:
Table of Contents

Advertisement

Splash Guard and Drip Tray
Figure 16
Caution
When installing the Racks, use the Splash Guard. Lower the Racks slowly.
[6] Install the Racks, the Crossover Assemblies, and the Dryer Assembly. Check that each Assembly seats firmly.
[7] Turn on the main power, the water, and the Processor.
[8] Check the solutions. If the surfaces of the solutions are moving, the Recirculation Pumps are operating.
[9] When the Temperature Ready Light flashes, use a thermometer of known accuracy and check the solution
temperature in the developer Tank.The correct temperature is:
7C8770 – 19JAN99
DRIP TRAY
33.3 0.3 °C (92.0 0.5 °F)
Standard Cycle
34.4 0.3 °C (94.0 0.5 °F)
Rapid Cycle