Archiving
A scanner and EchoPAC PC in a direct
connect environment
In this scenario the data is transferred from the Vivid i to a
dedicated EchoPAC PC workstation over the Ethernet (either in
a peer-to-peer connection with a crossover cable, or in a
network). The database from the EchoPAC PC is used as the
master and images are stored directly to the EchoPAC PC
internal harddrive. In this configuration the scanner is just an
intermediate acquisition unit which after completion of a study,
will not contain any patient information, measurements or
images.
Up to three scanners can be connected to one EchoPAC PC if
the workstation has the EchoPAC Share option enabled.
Vivid i: dataflow RemoteArch-RemoteHD
EchoPAC PC: dataflow LocalArchive-Int.HD
Figure 8-30: A scanner and EchoPAC PC in a direct connect
environment (peer to peer or network)
The acquisition can be done online or offline. Both situations
are described below.
Scanner's connectivity configuration
If working in a peer-to-peer connection with a crossover cable
between a Vivid i and an EchoPac PC, the default delivery
TCP/IP settings should be used.
If working in a network, follow the procedure described below to
configure the scanner.
192

Vivid i User's Manual

2378958-100 Rev. 02