Contrast Imaging
Frame rate
Lower frame rate
Controls the line density.
gives better resolu-
Up/Down
tion.
Enables the 2D image to be flipped 180 degrees.
Left/Right
Enables the display of a mirrored image. When applied, the
reference marker V moves to the other side of the image.
Angio
Turns Coded harmonic angio (color) on/off.
B Color maps
Displays a 2D maps menu to optimized the grey scale
presentation. The menu enables an option from a list of
non-linear grey-curves or different 2D-colorized curves to be
selected.
LV Contrast Soft menu controls
Power
Controls the amount of acoustic power applied to the
transmitted pulse.
Too high Power lev-
Compress
el will destroy the
Controls the degree of image contrast.
contrast agent.
Reject
Controls the Echo rejection level. When increased, low level
echoes are rejected and appear darker in the 2D image.
Dynamic Range
Controls the image contrast. A high dynamic range setting
gives a softer image.
Tilt
Enables the axis of the 2D image to be tilted to the left or to the
right. By default the axis of the 2D image is vertical.
Contour
Controls the image processing related to the extent of edge
enhancement applied.
106