Replace The Power Inlet Fuse - ABB ICOS GLA431-MCIA User Manual

Enhanced performance benchtop continuous flow analyzer
Table of Contents

Advertisement

USER MANUAL | ICOS | INSTRUCTIONS | UM/ICOS-EN REV. B.2

Replace the Power Inlet Fuse

If the fuse on the power inlet blows or is otherwise damaged, the analyzer shuts down. To
replace the fuse:
1.
Unplug the analyzer.
2. On the back panel of the analyzer, locate the fuse above the power inlet. (Figure 47)
3. Use a flathead screwdriver to remove the fuse.
a. Insert the head of the screwdriver into the slot below the fuse. (Figure 48)
b. Push down on the screwdriver handle to remove the fuse holder from the power
inlet.
c.
Remove the fuse from the fuse holder. (Figure 49)
Figure 49: Remove the Fuse from the Fuse Holder
4. Insert a new fuse into the fuse holder.
5. Re-insert the holder into the power inlet. Push it in until you hear a click.
6. Plug the power cord into the back panel of the analyzer.
7. Resume analyzer operation.
Figure 47: Analyzer Fused Inlet
Figure 48: Remove the Fuse
61

Advertisement

Table of Contents