Connect the Pretreatment Box, Pump, and Analyzer

1.
Connect the external pump to the pretreatment box TO EXT. PUMP port using the 6'
x 3/8" Swagelok connector and tubing (provided).
2. Connect the external pump's power cord from the pump to the EXT. PUMP POWER
outlet on the analyzer.
3. Connect the pretreatment box from the ANALYZER EXIT port to the analyzer's TO
EXT. PUMP port using the 6' x 3/8" Swagelok connector and tubing in the pump
connection kit (provided).
4. Connect the pretreatment box ANALYZER INLET port to the INLET port on the
analyzer, using the 6' x 1/4" Swagelok connector and tubing in the pump connection
kit (provided).
Figure 13 shows the pretreatment box plumbing diagram.
Figure 13: Pretreatment Box Plumbing Diagram