ABB ICOS GLA431-MCIA User Manual page 73

Enhanced performance benchtop continuous flow analyzer
Table of Contents

Advertisement

USER MANUAL | ICOS | INSTRUCTIONS | UM/ICOS-EN REV. B.2
b. Scroll to New port forward and set VNC: (Figure 70)
i.
Type Name: VNC
ii.
Set External port to 5900
iii.
Set Internal port to 5900
iv.
Select the Internal IP address drop down box and select the
analyzer's IP address.
v.
Press Add
c.
Set SSH: (Figure 71)
i. Type Name: SSH
ii. Set External port to 22
iii. Set Internal port to 22
iv. Select the Internal IP address drop down box and select the
analyzer's IP address.
v. Press Add
d. Set MODBUS: (Figure 72)
vi. Type Name: MODBUS
vii. Set External port to 2222
viii. Set Internal port to 2222
ix. Select the Internal IP address drop down box and select the
analyzer's IP address.
x. Press Add
Figure 70: New port forward - VNC
Figure 71: SSH
Figure 72: MODBUS
73