USER MANUAL | ICOS | INSTRUCTIONS | UM/ICOS-EN REV. B.2
8. Each step is displayed in the lower-right panel of the calibration screen as the
analyzer performs the calibration. Figure 38 shows the calibration process as a flow
chart.
9. When the Calibration Complete message is displayed, click the CLOSE button.
10. Enter TimeChart by selecting the Display button on the User Interface Control Bar ,
and verify that the displayed concentration correctly corresponds to your local gas
standard.
11. Repeat steps 1-10 for each gas you wish to calibrate.
The time of latest calibration is stored in Reference Gas Settings within
the Calibration menu for future reference.
Figure 38: Calibration Flow