Batch Mode Processing
In batch processing mode, the analyzer:
1.
Initiates the batch injection procedure.
2. Evacuates the cavity.
3. Flushes the cavity with zero air twice before requesting the sample.
4. Prompts you to inject sample (of >60 ml) gas into the syringe port.
a. You have 120 seconds to inject the sample.
If you take longer than 120 seconds to complete the injection, the Failed
Injection message displays and instructs you to restart the injection
process.
Figure 118 shows the Batch Injection Measurement screen displayed as a flow diagram to
show the batch measurement procedure.
Figure 118: Batch Injection Flow