USER MANUAL | ICOS | INSTRUCTIONS | UM/ICOS-EN REV. B.2
Service
ABB-trained field service engineers monitor the performance of the analyzer via the Service
Figure 43
screen. (
)
These settings determine the level of change that could affect measurement
•
performance.
The alarm threshold levels are analyzer-dependent and are set based upon the last
•
fixed setting.
The Service tab contains 3 Service buttons:
Reset Tau/Power button – resets the mirror ringdown time and laser power to
•
current settings. This is typically done after mirrors have been cleaned.
Reset Counter button – resets the # of days that maintenance is required. This is
•
typically done after yearly maintenance.
Launch TeamViewer button – TeamViewer allows service engineers to remotely
•
access the analyzer if service needs are required.
Figure 43: Service Screen
58