Step 3. Optional Step
Press the
the numeric keys and the Alpha Editor.
Step 4. Press the
Step 5. Enter the first frequency using the numeric keys in the table using the
unit termination menu keys.
Step 6. Press the
Tab —>
enter the corresponding ENR value of the ENR list.
When terminating the ENR value you can use either
keys. However, the result which appears in the table is in dB.
Step 7. Press the
enter the next frequency value on the ENR list.
Step 8. Repeat steps 5 to 7 and until all the frequency and ENR values you need
are entered.
Step 9. After completing the ENR table entries, press the
return to the ENR menu.
Step 10. Once you have completed entering the ENR data, save the ENR table
using the
Chapter 2
menu key and enter the noise source model number using
ID
menu key to enter the noise source ENR values.
Edit Table
key to move the highlight to the ENR Value column and
key to move the highlight to the Frequency column and
Tab —>
key.
File
Making Basic Measurements
Entering ENR Data
,
,
, or
dB
K
C
F
key or
Prev
ENR
menu
key to
25