Report Options (continued)
Option
Smooth diff?
Molecular area
Density conversion Enter the adsorptive gas-to-liquid density conversion factor.
Nonideality
Gemini VII Keypad Operator Manual
239-42801-01 (Rev E) — July 2021
Description
Smooths the differential calculation, eliminating variations in the
differential computation caused by noise in the input data.
Choices: Yes, No
Enter the molecular cross-sectional area of the adsorptive gas.
Range: 0.001 to 1.000 nm2
Range: 0.0000001 to 1.0000000
Nonideality is calculated when a free-space value (other than none) is
selected. Enter the nonideality correction factor for the adsorptive gas.
Range: 0.0 to 99.9 %/atm
Report Options