3 Command Prompts
Report Options (continued)
Option
Area points to
Report BET multi-pt Choices: Yes, No, Pass/Fail
3 - 22
Description
Enter the relative pressure at which to stop selecting collected data
points for calculating surface area. If the range specified includes more
than 50 data points, only the first 50 points are used to calculate surface
area. Individual points after analysis may be selected or deselected.
Range: Area points from value to 1.0000 P/Po
If Pass/Fail is selected, specify a range.
Range: 0.0000 to 9999.9999 m2/g (for both prompts)
Gemini VII Keypad Operator Manual
239-42801-01 (Rev E) — July 2021
Need help?
Do you have a question about the GEMINI VII KEYPAD and is the answer not in the manual?
Questions and answers