Micromeritics GEMINI VII KEYPAD Operator's Manual page 70

Surface area analyzer
Table of Contents

Advertisement

3 Command Prompts
Analysis Condition Prompts (continued)
Field or Button
Equilibrate
Equilibration time
3 - 18
Description
Selects the analysis mode.
Choices: Equilibrate, Scan
Increases the pressure in steps to the next specified level for data
collection. It maintains this pressure and monitors the volume
adsorbed. When the rate of adsorption (volume adsorbed during the
equilibration time divided by total volume adsorbed) falls below
0.01% and the variance from 11 consecutive readings falls below
0.1%, equilibration is assumed to have occurred. The analyzer then
introduces the next adsorbate gas dose.
Enter the equilibration time.
Range: 1 to 1000 sec
Gemini VII Keypad Operator Manual
239-42801-01 (Rev E) — July 2021

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the GEMINI VII KEYPAD and is the answer not in the manual?

Table of Contents