3 Command Prompts
Analysis Condition Prompts (continued)
Field or Button
Pressure table
Replace
3 - 16
Description
Selects the means for changing or preserving the currently stored tar-
get relative pressure table.
Choices are: Replace, Preserve, Edit
The highest adsorption point prior to saturation should
be 0.9990 P/Po. If saturation pressure has decreased
significantly since the previous measurement, target
pressures close to 1.0 P/Po can result in premature sat-
uration.
Replaces the currently stored pressure table with a new table of
evenly spaced points.
Enter the First rel. pressure for data collection.
Range: 0.00001 to 1.0000 P/Po
Enter the Last rel. pressure for data collections.
Range: First rel. pressure to 1.0000 P/Po
Enter the same value entered for First rel. pressure if performing a
single-point analysis. The analyzer collects the specified number of
data points at evenly spaced pressure intervals during the period of
analysis. Enter 1 if performing a single-point analysis.
Gemini VII Keypad Operator Manual
239-42801-01 (Rev E) — July 2021