Preparing Vscan Air CL for Use
Environmental requirements for Vscan Air CL
Description
Temperature
0°C to + 35°C
Humidity
10–90%
Air pressure
54 kPa to 106 kPa
Transient operating conditions
NOTE:
NOTE:
3-8
Environmental requirements
Table 3-1:
Environmental requirements
Operational
Permissible transient environmental operating conditions:
•
Temperatures ranging from -20°C to 0°C or from +35°C to
+50°C.
•
Device will function for a minimum of 20 minutes when
placed in an environment with temperatures ranging from
-20°C to 0°C or from +35°C to +50°C after storage at room
temperature (20 ± 2°C).
•
Following storage at temperatures ranging from -20°C to
0°C or from +35°C to +50°C, the device will within 10
minutes after being returned to room temperature (20 ±
2°C), function for a minimum of 20 minutes.
Avoid exposing the unit to saline moisture. In case of exposure
to saline moisture, clean the unit as described in 'Cleaning and
disinfection' on page 6-6.
Image on the display device hosting the Vscan Air app is
dependent on ambient light; avoid direct sunlight or reflections
from other light sources on the display when scanning and
reviewing images. The display viewing angle should be as small
as possible.
If you are having difficulty seeing the image due to the lighting
conditions, try to change brightness of the display device or
change your position/location of use.
Nonoperational
- 40°C to + 70°C
10–95%
50 kPa to 106 kPa
Direction GP092020-1EN Rev 18
Storage and transport
- 40°C to + 70°C
10–95%
50 kPa to 106 kPa
–
Vscan Air
User Manual