System Care And Maintenance; Overview - GE Vscan Air Manual

Table of Contents

Advertisement

Vscan Air Maintenance

System care and maintenance

Overview

CAUTION
NOTE:
6-2
Only trained persons should perform the safety inspections.
The Vscan Air CL requires regular care and maintenance to
function safely and properly.
The expected service life of the Vscan Air CL is 5 years.
To ensure that the Vscan Air constantly operates at maximum
efficiency, we recommend that the following procedures be
observed as part of the customer's internal routine maintenance
program.
The user must ensure that safety inspections are performed at
least every 12 months according to the requirements of the
patient safety standard EN/ES/IEC 60601-1 and its national
deviations or according to the requirements of the IEC62353
standard, medical electrical equipment – recurrent test and test
after repair of medical electrical equipment.
Special attention shall be paid to the following:
Safety related marking / labels are legible and complete for
the Vscan Air probe
The integrity of the probe e.g. is there any observed
damage to the Vscan Air probe
Assess the accessories for damage (AC adapter, USB
cable, wireless charger pad), and
Assess if the required documentation is available and
reflects the current revision and/or configuration of the
Vscan Air
Vscan Air
Direction GP092020-1EN Rev 18
User Manual