Vscan Air Maintenance
Quality assurance
Typical Tests to Perform
6-4
A good Quality Assurance Evaluation program consists of
periodic systematic actions that provide the user with adequate
confidence that their diagnostic ultrasound system will produce
consistently high-quality images and quantitative information.
Therefore, it is in the best interests of every ultrasound user to
routinely monitor equipment performance.
Quality assurance measurements provide results relating to
system performance. Typically. these are:
Axial Measurement Accuracy
Lateral Measurement Accuracy
Axial and Lateral Resolution
Penetration
Functional and Contrast Resolution
Gray Scale Photography.
With these tests, a performance baseline can be set at
installation with the phantom in your department. Future test
results can be compared to the baseline in order to maintain a
record of system performance trends.
Vscan Air
Direction GP092020-1EN Rev 18