Acclimatization time
Other environment: Aircraft Ambulance/Road Ambulance
Vscan Air
User Manual
Direction GP092020-1EN Rev 18
Allow the Vscan Air CL to acclimate for approximately 10
minutes, if stored at temperatures ranging from -20°C to 0°C or
in temperatures ranging from +35°C to +50°C.
Allow the Vscan Air CL to acclimate for approximately 30
minutes, if stored at temperatures ranging from -40°C to -20°C
or in temperatures ranging from +50°C to +70°C.
The Vscan Air ultrasound system can be used in an emergency
medical services environment, including road** and air
ambulances*.
The Vscan Air CL is not certified for being mounted or fixed
inside a road- or air ambulance.
The Vscan Air CL can be used in emergency medical
services environments. See 'Environmental requirements
for Vscan Air CL' on page 3-8 for more information.
The Vscan Air CL probe is not certified for being charged
inside a road- or air ambulance.
The AC Adapter must not be operated above 5000m.
* The Vscan Air ultrasound system is compliant to IEC
60601-1-12 and EN13718 as stated in 'Regulatory
requirements' on page i-4. Additional regulations might apply
** The Vscan Air ultrasound system is compliant to IEC
60601-1-12 and EN1789 as stated in 'Regulatory requirements'
on page i-4. Additional regulations might apply.