Figure 2.6-33: Drift and Velocity Tach Calibration
2.6.7.5

Additional steps for Auto-Traverse and Comprehensive chairs

For otolith function testing, a 0 to 7 cm micro-centrifuge option is available. The chair seat should be
centered laterally on the chair base. If the scale shows the chair is off-center, enter the current chair
off-center position under "Current Lateral Position (cm)", then select the direction the chair is from center
(Left / Right). Click on the Center Laterally button to move the chair to center.
Caution: Do not attempt to move the chair frame laterally by pushing on the chair frame or damage may
occur.
Figure 2.6-34: Additional Off-Axis Center settings for Auto-Traverse and Comprehensive chairs
D-0134090-A – 2022/11
VisualEyes™ - Instructions for Use - EN