Guidance and Manufacturer's Declaration — Electromagnetic Immunity
The ORION is intended for use in the electromagnetic environment specified below. The customer or the
user of the ORION should assure that it is used in such an environment,
Immunity test
IEC / EN 60601 test
level
Conducted RF
3 Vrms
IEC / EN 61000-4-6
150kHz to 80 MHz
6 Vrms in ISM bands
150kHz to 80 MHz
80 % AM at 1 kHz
Radiated RF
3 V/m
IEC / EN 61000-4-3
80 MHz to 2,7 GHz
80 % AM at 1 kHz
D-0134090-A – 2022/11
VisualEyes™ - Instructions for Use - EN
Compliance level
Electromagnetic environment –
guidance
Portable and mobile RF
communications equipment should
be used no closer to any parts of
the ORION, including cables, than
the recommended separation
distance calculated from the
equation applicable to the
frequency of the transmitter.
Recommended separation
distance:
3 Vrms
6 Vrms
3 V/m
Where P is the maximum output
power rating of the transmitter in
watts (W) according to the
transmitter manufacturer and d is
the recommended separation
distance in meters (m).
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey,
should be less than the compliance
level in each frequency range.
Interference may occur in the
vicinity of equipment marked with
the following symbol:
80 MHz to 800 MHz
800 MHz to 2,7 GHz
a
b