Filtration
7.2
Part II (optional):
Emptying with Pressure After Filtration
If not needed proceed to Part III – Disconnection.
Option A: Emptying using inlet line
1. Close the inlet valve (8) and venting line valve (10). Possibly not needed
if other equipment (e.g. Maxicaps
inlet side and the pressurized air supply is also connected to this
equipment so that the emptying is performed at once for the complete
equipment-chain.
2. Connect the supply of pressurized air to the Inlet (3).
3. Apply max. pressure of 0.5 bar | 7.2 psi.
4. Empty the system with pressurized air in the direction of the outlet of the
device.
5. Stop the air supply after emptying, close the inlet valve (8) and outlet valve
(7) and open venting line valve (10).
6. Wait until the pressure is released through the opened air filter (1).
7. Close the opened venting line valve (10), when the device is not under
pressure anymore.
Option B: Emptying using the air filter .
1. Close the inlet valve (8).
2. Connect the pressurized air supply to the Tri-Clamp of the air filter (1).
3. Open venting line valve (10), apply max. pressure of 0.5 bar | 7.2 psi.
4. Empty the system with pressurized air in the direction of the outlet of the
device.
5. Stop the air supply after emptying, close venting line valve (10) and outlet
valve (7).
6. Disconnect the pressurized air supply.
7. Open the venting line valve (10).
8. Wait until the pressure lowers through the air filter.
9. Close the venting line valve (10).
30
Instructions for Use Maxicaps
MR Virosart
Max) is connected to the
®
®
MR Virosart
HF, Max and Media
®
®