Front-Panel Key Reference
Source
The output power of the tracking generator is swept according to the
sweep rate of the analyzer. The output power is always swept from the
source power setting to a higher power setting (negative source power
sweep values are not allowed).
Power-sweep measurements are particularly useful in making gain
compression measurements or output power versus frequency
measurements.
Key Access:
Allows you to select between automatic and manual adjustment of the
Attenuation
tracking generator's output attenuator. The Agilent E7401A can be
Auto Man
manually adjusted from 0 to 60 dB in 10 dB steps. All other Agilent
EMC analyzers can be manually adjusted from 0 to 56 dB in 8 dB steps.
When auto-coupled, the attenuation function automatically adjusts the
attenuator to yield the source amplitude level specified by the
On Off
for power sweeps greater than 10 dB.
Key Access:
When the source attenuation is set manually, the actual source
CAUTION
amplitude may be greater than the amplitude indicated by the source
amplitude. With source attenuation set to manual, source amplitude is
permitted to be set to values beyond output levels to accommodate the
full range of the analyzer capabilities. As a result, exercise caution
when connecting a power-level sensitive device to the tracking
generator output
Allows you to set the step size of the power level range of the tracking
Amptd Step
generator. The default setting is one vertical scale division when in a
Auto Man
logarithmic amplitude scale.
Key Access:
Offsets the displayed power of the tracking generator. Using the
Amptd Offset
amplitude offset capability of the tracking generator allows you to take
system losses or gains into account, thereby displaying the actual power
delivered to the device under test.
Key Access:
Key Access:
Normalize
Store Ref
(1→ 3)
182
Source
softkey function. For the Agilent E7401A, press
Source
Source
Source
Source
.
Copies trace 1 into trace 3.
pressed before pressing
is pressed before
Store reference trace before turning on
Normalize is displayed in the status line, however
Normalize has been activated.
Store Ref (1→ 3)
Normalize (On)
, the error message:
Store Ref (1→ 3)
Amplitude
Attenuation (Man)
must be
. If
Normalize (On)
Chapter 2