Agilent Technologies EMC Series User Manual page 130

9 khz- 1.5/3.0/6.7/13.2/26.5 ghz
Hide thumbs Also See for EMC Series:
Table of Contents

Advertisement

Front-Panel Key Reference
Meas Setup
Interpolation modes determine how limit values are computed between
NOTE
points in the limit table. The appearance of a limit trace is also affected
by the amplitude scale, which may be linear or logarithmic.
Selects whether limit lines will be entered using frequency or sweep
X Axis Units
time to define the segments. They can be specified as a table of limit
Freq Time
line segments of amplitude versus frequency, or of amplitude versus
time. Time values are evaluated with respect to the analyzer sweep
time. A time value of zero corresponds to the start of the sweep, which
is at the left edge of the graticule.
Switching the limit line definition between frequency and time will
erase both of the current limit lines. The message Changing X axis
units will delete all limits. If you are sure, press key
again to change units will appear. Press
to purge both limit lines and switch between frequency and time.
Key Access:
Allows you to choose fixed or relative limit lines. The fixed (
Limits
uses the current limit line as a reference with fixed frequency and
Fixed Rel
amplitude values. The relative (
line value to be relative to the displayed center frequency and reference
level amplitude values. When limit lines are specified with time, rather
than frequency, the
130
frequency interpolation is logarithmic,
frequency values between limit points
are computed by first taking the
logarithm of both the table values and
the intermediate value. A linear
interpolation is then performed in this
logarithmic frequency space. An exactly
analogous manipulation is done for
logarithmic amplitude interpolation.
Key Access:
Limit 1
Amptd Interp
Allows you to determine how limit
Log Lin
trace values are computed between
points in a limit table. The available
interpolation modes are linear and
logarithmic. If the linear mode is used
for both frequency and amplitude, a
straight line is used when interpolating
between points in a limit table.
Key Access:
or
,
Meas Setup, More
Limits
Rel
setting only affects the amplitude values. The
Rel
Meas Setup, More
or
,
Limit 2
More
Meas Setup
,
Limit 2
More
X Axis Units Freq Time
) setting causes the current limit
,
,
Limits
,
,
Limits
Limit 1
again
) type
Fixed
Chapter 2

Advertisement

Table of Contents
loading

This manual is also suitable for:

E7401aE7402aE7403aE7404aE7405a

Table of Contents