Front-Panel Key Reference
Meas Setup
This key sets the analyzer up for making CISPR BAND D
measurements. The analyzer settings for this setup are as follows:
Description
Start Frequency
Stop Frequency
Resolution Bandwidth
Video Bandwidth
Sweep Time
Reference Level
Attenuation
Detector
Scale Type
Scale/Division
Marker 1
Key Access:
Accesses menu keys that allow you to create and define the parameters
Limits
of limit lines.
Key Access:
Limit 1
Limit 2
126
Setup
200 MHz
1 GHz
120 kHz
300 kHz
Autocoupled
80 dBµV
5 dB
Peak
Log
10 dB/Div
Normal, active
Meas Setup
Meas Setup, More
Allows you to select either of the two available limits.
The
and
Limit 1
Limit 2
you to set parameters for the selected limit.
Key Access:
Meas Setup, More
Type
Allows you to define the limit you are
Upper Lower
editing as either an upper or lower
limit. An upper limit fails if the trace
exceeds the limit. A lower limit fails if
the trace falls below the limit.
Key Access:
Limit 1
Turns the limit line display on and off.
Limit On Off
Either
must be turned on for a limit test to
work.
keys access menu keys that allow
,
Limits
Meas Setup, More
or
Limit 2
or
, as well as
Limit
Margin
,
,
Limits
,
Test
Chapter 2